Application: |
Measurement of beer, in process lines. |
Concentration measurement:
|
0...18.5Brix for “REGULAR” Scale; 0.1% of the scale range, maximum accuracy ±0.02Brix with ±10°C (±18°F) variation.
0...2000mBrix for “DIET” Scale; 0.15% of the scale range, maximum Accuracy ±0.003Brix with ±2.5°C (±3.6°F) |
CO 2 measurement :
|
0...5 v/v (0...10g/l); ±0.05 v/v (±0.1g/l) |
| ALCOHOL measurement: |
Alcohol: 0...10% w/w; ±0.02 w/w
Real Extracts: 0...18.5 Ew; ±0.02 Ew
Original Extracts: 0…18.5 °Plato; ±0.02 °Plato |
OXYGEN measurement (optional) :
|
0…10 ppm; ±0.01 ppm |
pH measurement (optional) :
|
2…10 pH; ± 0.05 pH |
| Product temperature: |
-5°C...+35°C [+23°F…+95°F]
(-5°C…+90°C [+23 °F…+194 °F] for “Regular” scale sugared drinks) with automatic compensation of sugar concentration by means of a Pt1000 temperature probe. |
Power Supply:
|
AC 3/PE 230/400V ±10% 50Hz ±2% 1000VA
AC 3/PE 460V ±10% 60Hz ±2% 1000VA |